Introduction
The term osseointegration is defined as a direct structural and functional linkage between living bone tissue and the implant surface that is under loading without fibrous tissue. Osseointegrated dental implant-supported prostheses have become a scientifically agreeable and frequently used choice in the treatment of complete and partial edentulism involving the restoration of lost teeth.1,2
In addition to place parameters of dental implant material, bone tissue quality and quantity are very significant agents for accomplished osseointegration. The length of time between surgery and loading and the nominal clinical success percentage in poor quality bone tissues are the most challenging problems in dental implant treatments.3,4
In order to prevent inadequate bone tissue quality and amount, various techniques for dental implant treatment have been used to improve bone tissue creation and osseointegration. For example, tissue growth factors, melatonin hormone and antiosteoclastic drugs, and bisphosphonates have been used to improve the bone integration of titanium implants.5,6
Success in dental implants depends on osseointegration. Accordingly, research on osseointegration in implants has lately intensified on the additional use of drugs known to affect bone turnover.7 In particular, in animal fracture models and clinical studies, bisphosphonate class drugs have been successfully applied to improve healing and osseointegration.8
Bisphosphonates are drugs that have a high affinity for bone hydroxyapatite. Therefore, its pharmacological effects on bone play a role in bone reconstruction or destruction. We can group bisphosphonates as first generation nitrogen-free (clodronate, etidronate and tiludronate drugs) and second and third generation nitrogen-containing (alendronate, risedronate, ibandronate and zoledronate drugs). Thanks to oral bisphosphonates are poorly absorbed, the percentage bioavailability is lower than intravenous bisphosphonates. Oral bisphosphonates contain alendronate, risedronate, etidronate, and tiludronate. Pamidronate and zoledronate are administered only intravenously, while ibandronate and clodronate are administered by both routes.9
Bisphosphonates, widely used in the prevention of metastatic bone cancers, Paget’s disease, osteoporosis, multiple myeloma, hypercalcemia, osteogenesis imperfecta and other metabolic diseases of the bone, work by suppressing osteoclast cells that destroy bone. In particular, zoledronic acid (ZOL), the most clinically potent bisphosphonate, is frequently used in the theraphy of osteoporosis disease. Bisphosphonates have been reported to increase the differentiation, proliferation and maturation of osteoblasts.10 It has been reported that bisphosphonates have antiosteoclastic and proosteoblastic effects on the bone tissue repair system. Although diverse speculations have been proposed, the system of interaction between bisphosphonates and bone tissue has not yet been completely illuminated. According to past studies, bisphosphonates immediately limit the activation of osteoclast cells and may induce mineralized bone formation through osteoblast cells and thus reduce bone loss.4,11,12
Surface features of dental implant material are among the significant parameters affecting bone implant linkage. In clinical research in which implant surfaces were evaluated in terms of bone implant fusion, it was observed that no implant surface system was completely superior to any other.13,14
The reason why dental implants can be considered successful in the long term lies in the surface features of the implants. While the surface features of dental implants were machine turned surfaces of pure titanium in the 1970s, they have now changed to a microroughened surface. Although the techniques we have just mentioned are used in the surface treatment of dental implants, there is currently the development of new biomedical surfaces. After the 1990s, nano touch for the regulation of dental implant surfaces has been examined. Additionally, surface nanotopography and chemistry have been reported to influence protein adsorption, osteogenic cell behavior, and bone implant interaction.15
In order for changes on dental implant surfaces to be positive, the nature of osseointegration must be understood. The term osseointegration is thought to be a type of bone healing accompanied by an inflammatory response.16 In order to improve osseointegration, the nanotopography, surface chemistry and surface energy of the dental implants have been changed. However, these changes also altered in vitro cellular behavior and in vivo bone physiology.17,18
In our study, we aimed to examine the healing physiology of zoledronic acid application and the interactions between bone and dental implants by including dental implant surface modifications designed to improve the healing between the bone and the implant surface. The purpose of our research was to mechanically appraise the effect of local ZOL application on implant bone fusion in implants with different surface properties.
Material and Methods
Animals and Study Design
Approval for the designed experimental setup and study was granted by the Firat University Animal Experiments Local Ethics Committee (02.02.2021, approval no.: 2021/02-449). Rats were provided by the Firat University Experimental Research Center (Elazığ, Türkiye). This study was conducted in Elazığ, Turkey, within the scope of the animal experiment protocol of the Ministry of Agriculture of the Republic of Turkey, led by the recommendations of the European Declaration of Helsinki. In the studies conducted, the experiments were carried out without causing any pain to the animals, and all ethical rules were followed. Hole animals used in the experimental stage were obtained from the same center. Advice in the Helsinki Declaration regarding the protection of laboratory experimental animals was followed.
The experimental setup was conducted using 60 female Sprague Dawley rats aged 6–12 months. At the beginning of the empirical study, the average body weight of the female rats was determined as 270–300 g. The rats were housed in special rooms where the temperature was continuously controlled and fed ad libitum, with a 12 hour light and 12 hour dark cycle. As a result of the statistical analyses made before the experiments, it was determined that the number of rats in each group should be at least eight. Considering that rats might die during surgical procedures and experimental setup, the study started with 10 rats in each group. A vaginal smear was taken for standardization, and all rats were in the same oestrus period. Sixty (60) female rats were randomly divided into six equal parts.
Machine-surfaced control group (MAC-C) (n = 10): Machine-surfaced titanium implants were placed into the corticocancellous bone of the right tibias of the rats. After a four-week healing period, the rats were sacrificed.
Resorbable blast media (RBM) surfaced control group (RBM-CNT) (n = 10): RBM-surfaced titanium implants (average roughness (Ra): 1–2) were placed into the corticocancellous bone of the right tibias of the rats. After a four-week healing period, the rats were sacrificed.
Sandblasted, large grit, acid etched (SLA) surfaced control group (SLA-CNT) (n = 10): SLA- surfaced titanium implants (average roughness (Ra): 1–2) were placed into the corticocancellous bone of right tibias of the rats. After a four-week healing process, the rats were sacrificed.
Machine-surfaced local ZOL group (MAC-LZA) (n = 10): The machine-surfaced titanium implants were placed into the corticocancellous bone of the right tibias of the rats by local application of 4 mg/5 ML of zoledronic acid to the implant socket, and, after a four-week healing process, the rats were sacrificed.
RBM-surfaced local ZOL group (RBM-LZA) (n = 10): The RBM-surfaced titanium implants (average roughness (Ra): 1–2) were placed into the corticocancellous bone of the right tibias of the rats by local application of 4 mg/5 ML of zoledronic acid to the implant socket, and, after a four week healing process, the rats were sacrificed.
SLA-surfaced local ZOL group (SLA-LZA) (n = 10): The SLA-surfaced titanium implants (average roughness (Ra): 1–2) were placed into the corticocancellous bone of the right tibias of the rats by local application of 4 mg/5 ML of ZOL to the implant socket, and, after a four-week healing process, the rats were sacrificed.
Surgical Procedures
Hole surgical interventions were performed under general anaesthesia in sterile conditions. General anaesthetics (10 mg/kg xylazine and 40 mg/kg ketamine) were administered intraperitoneally to the rats. After the surgical area was shaved, it was cleaned with povidone-iodine solution. During the surgical placement of the implants, a 2 cm incision was made over the right tibia bones of the rats with a scalpel (no. 15), and the metaphyseal part of the bone was reached. The surgical area was cleaned with sterile saline solution to prevent heating while the implant sockets were opened. After 4 mm long and 2.5 mm diameter implants (TiAl6V4, Implance Dental Implant Systems, AGS Medikal Corporation, Istanbul, Turkey) were placed into the corticocancellous bone in the right tibias of the rats, the soft tissues were sutured with 4–0 absorbable sutures (polyglactin) (Figure 1). After the completion of all surgical procedures, antibiotics (40 mg/kg cephalosporin) and analgesics (0.1 mg/kg tramadol hydrochloride) were administered intramuscularly for three days to prevent infection and pain. The rats were sacrificed at the end of the four-week experimental period following the surgical interventions. Titanium implants were collected along with the surrounding bone tissue and fixed in formaldehyde solution. A biomechanical test device analysed the osseointegration of the obtained implant samples by applying reverse torque (Mark 10, NY, USA). Data are recorded as force (N/cm) (Figure 2).
Figure 1 Surgical placement of the implants in the corticocancellous bones of the metaphyseal parts of the right tibias of rats. |
Figure 2 Biomechanical; reverse torque analysis, of the implants (Mark-10 Cap Torque Tester, Model TT01-12, NY, USA). |
Statistical Analysis
IBM SPSS Statistics 22 program was used for statistical analysis of the data in the study. In this evaluation, the conformity of the parameters to the normal distribution was evaluated with the Kolmogorov–Smirnov test. In addition, the Kruskal–Wallis test was used to compare the non-normally distributed parameters between groups, and the Mann–Whitney U-test was used to determine the group that caused the difference. Significance was evaluated at the p
Results
Data analysis showed that the biomechanical bone implant fusion values of the MAC-LZA, SLA and RBM groups were higher than those of the MAC-C group (p 0.05) (Table 1).
Table 1 Biomechanical Bone Implant Connection (N/Cm) Levels of the Groups |
Discussion
Research has shown that the success of dental implants is linked to a patient’s medical history. While there is an increase in the number of implants applied to the elderly population, osteoporosis is commonly observed among these patients.19 Studies have reported the negative effects of osteoporosis on the jawbone. It is reported that bisphosphonates are commonly used in the treatment of osteoporosis.20–22
In in vivo studies, bone implant fusion has been emphasized in the use of bisphosphonate locally in treatments. Local use of bisphosphonates in bone implant fusion has been reported to have favourable results.23–26
In this research, the effect of bisphosphonates used in the therapy of osteoporosis on bone implant fusion was evaluated. Similar results were obtained, and it was found that bisphosphonate use positively affected implant osteointegration.
In their study, Im et al applied implants with RBM and SLA surface properties to the canine region of the maxilla. They obtained positive results in early stability measurements with a periotest in both implant types. However, in the measurements performed at the 12th week, better results were obtained in implants with RBM surface properties compared to implants with SLA surface properties. In the statistical analyses, torque values and bone implant fusion rates were evaluated and no statistically significant difference was found. As a result, good results were obtained in osteointegration in both implant surface types, and it has been reported that both implant surface types show healing properties compatible with bone.27
In studies that compared SLA surface and RBM surface implants with machine-formed surfaces, more favourable results were obtained in osteoblast activity and bone–implant contact in implants with SLA and RBM surface properties.28,29
In a study comparing implants with SLA, RBM and RBM acid-etching surface properties, no significant difference was found in reverse torque and implant-bone fusion, although the surface area of implants with an SLA surface was higher.30
In the present research, the osteointegration values of implants with machined, SLA and RBM surfaces were evaluated by measuring the reverse torque. Significant differences in reverse torque values were obtained in implants with SLA and RBM surface properties compared to machined-surface implants. As in other studies, no significant difference was obtained between SLA and RBM surfaces.
In a study by Duarte et al, the ovaries of rats were excised and treated with alendronate and estradiol. The osseointegration of dental implants was then evaluated. Positive results were obtained in implant osseointegration, and it was suggested that the negative effects of oestrogen deficiency after ovariectomy could be eliminated with bisphosphonates.31
In a study by Giro et al, bisphosphonate treatment was applied and extraction torque was investigated in subjects with oestrogen deficiency. It was reported that oestrogen deficiency negatively affected the extraction torque, whereas bisphosphonate use reduced this negative effect.32
In a study by Astrand et al, ZOL was applied locally to HA-coated implants in ovariectomized rats. During the healing period, no adverse trabecular distribution was observed in the peri-implant tissues. An improvement in fixation was even observed. While resorption occurs in osteoclasts, this was reduced and prevented by bisphosphonate treatment. Bisphosphonates penetrate the mineral structure in the bone structure and cause inactivation of osteoclasts in the resorption process.24
In one study, the use of ZOL as a bisphosphonate was evaluated using histomorphometry and microtomography (micro-CT). Positive effects on trabecular number, thickness and connectivity were observed. An increase in thrust force was observed as a result of ZOL use, and this increase was found to be 8.4 times higher than normal bone.33 ZOL is reportedly involved in stimulating apoptosis of osteoclasts and macrophages and has effects on osteoclast mediated bone resorption.34
Jakobsen et al applied alendronate locally on dogs and evaluated the results for 12 weeks. It was reported that an increase in bone implant fusion occurred in hydroxyapatite-coated titanium implants compared to the control group.35 Cuaira’n et al applied orthodontic mini implants to dogs, and then 50 mL phosphate buffered solutions containing 16 mg ZOL were applied locally to the implant socket.36 The researchers reported that ZOL application had a positive effect on bone implant fusion. Titanium implants were applied to rat maxillae by Abtahi et al, and positive effects on bone implant fusion were observed.37
In other study, Sokmen et al reported that both local and systemic application of zoledronic acid in implants with RBM surface placed without primary stabilization increased osseointegration levels in implants with and without primary stabilization compared to controls and that local application could be a successful and safe method.38
Similarly, in this study, reverse torque values were found to be higher in implants with local ZOL than in those without. The highest osteointegration value was obtained in implants with local SLA surfaces, while the lowest value was obtained in implants with machined surfaces. In the comparison between RBM with local ZOL and SLA surfaces with local ZOL, higher values were obtained in implants with SLA surfaces. This suggests that the use of local ZOL has a positive effect on the osteointegration value as the surface area increases.
In a research conducted by Abtahi et al, they reported that there was no statistical difference between bisphosphonate coated dental implants and non-bisphosphonate coated dental implants in the early healing period, but dental implants not coated with bisphosphonate suffered more marginal bone loss than the other group.39 Zuffetti et al stated that local application of bisphosphonate drugs positively affects the survival of dental implants.40
In their study, where Fiorillo et al evaluated the effects of bisphosphonate drugs on the health dental implants, they argued that it could start recent opportunities in orthopedic surgery through osteoporotic bone and tooth rehabilitation in patient groups using bisphosphonate drugs.41
In line with the literature reviewed, although bisphosphonate drug groups make important contributions to various medical and dental treatments, their effects on dental implant procedures require careful evaluation. In addition, local applications of bisphosphonates positively affect the success rates of dental implant osseointegration. Therefore, ongoing study on this topic is extremely important for improving protocols for patients who use bisphosphonates and the success of dental implant applications.
Conclusions
According to the limited results of the study, it can be stated that local application of ZOL may be an effective method to increase osseointegration in machined, RBM and SLA surface implants. In addition, local application of ZOL may be a more reliable method than systemic application. There is a need for further studies examining the local ZOL application and osseointegration mechanism, considering different implant surface structuring methods.
In our study, it was concluded that local ZOL application had a positive impact on the bone around the dental implant. In conclusion, with our study, we can say that local application of ZOL to dental implants is a step forward in better understanding and measuring the basic biological process.
Acknowledgment
The authors wish to thank the Implance Dental Implant System, AGS Medical Corporation, Istanbul, Turkey, for providing the implants.
Author Contributions
All authors made a significant contribution to the work reported, whether that is in the conception, study design, execution, acquisition of data, analysis and interpretation, or in all these areas; took part in drafting, revising or critically reviewing the article; gave final approval of the version to be published; have agreed on the journal to which the article has been submitted; and agree to be accountable for all aspects of the work.
Disclosure
The authors declare that they have no competing interests in this work.
References
1. Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous Anchorage of dental prostheses I Experimental studies Scand.J Plast Reconstr Surg. 1969;3:81–100.
2. Buser D, Mericske-Stern R, Bernard JP, et al. Long-term evaluation of non-submerged ITI implants. Part 1: 8-year life table analysis of a prospective multi-center study with 2359 implants. Clin Oral Implants Res. 1997;8:161–172. doi:10.1034/j.1600-0501.1997.080302.x
3. Dursun CK, Dursun E, Eratalay K, et al. Effect of porous titanium granules on bone regeneration and primary stability in maxillary sinus: a human clinical, histomorphometric, and microcomputed tomography analyses. J Craniofac Surg. 2016;27:391–397. doi:10.1097/SCS.0000000000002421
4. Yaman F, Agacayak S, Atilgan S, et al. Effects of systemic zoledronic acid administration on osseointegration of hydroxyapatite-coated and resorbable blast material surface implants in rabbit models. Int J Oral Maxillofac Implants. 2012;27:1443–1447.
5. Dursun E, Keceli HG, Uysal S, et al. Management of limited vertical bone height in the posterior mandible: short dental implants versus nerve lateralization with standard length implants. J Craniofac Surg. 2016;27:578–585. doi:10.1097/SCS.0000000000002459
6. Dundar S, Yaman F, Saybak A, et al. Evaluation of effects of topical melatonin application on osseointegration of dental implant: an experimental study. J Oral Implantol. 2016;42:386–389. doi:10.1563/aaid-joi-D-16-00048
7. Choi JY, Lee H-J, Jang J-U, et al. Comparison between bioactive fluoride modified and bioinert anodically oxidized implant surfaces in early bone response using rabbit tibia model. Implant Dent. 2012;21(2):124–128. doi:10.1097/ID.0b013e318249f283
8. Tsetsenekou E, Papadopoulos T, Kalyvas D, et al. The influence of alendronate on osseointegration of nanotreated dental implants in New Zealand rabbits. Clin Oral Implants Res. 2012;23(6):659–666. doi:10.1111/j.1600-0501.2011.02189.x
9. de-Freitas NR, Lima LB, de-Moura MB, Veloso-Guedes CC, Simamoto-Júnior PC, de-Magalhães D. Bisphosphonate treatment and dental implants: a systematic review. Med Oral Patol Oral Cir Bucal. 2016;21(5):e644–e651. doi:10.4317/medoral.20920
10. Lee SY, Koak JY, Heo SJ, et al. Osseointegration of anodized titanium implants coated with poly(lactide-co-glycolide)/basic fibroblast growth factor by electrospray. Int J Oral Maxillofac Implants. 2010;25(2):315–320.
11. Little DG, Comell MS, Briody J, et al. Intravenous pamidronate reduces osteoporosis and improves formation of the regenerate during distraction osteogenesis. J Bone Joint Surg Br. 2001;83:1069–1074. doi:10.1302/0301-620X.83B7.0831069
12. Dhillon S. Zoledronic acid: a review in osteoporosis. Drugs. 2016;76:1683–1697. doi:10.1007/s40265-016-0662-4
13. Esposito M, Ardebili Y, Worthington HV. Interventions for replacing missing teeth: different types of dental implants. Cochrane Database Syst Rev. 2014;7:CD003815.
14. Jokstad A, Braegger U, Brunski JB, Carr AB, Naert I, Wennerberg A. Quality of dental implants. Int Dent J. 2003;53(6 Suppl 2):409–443. doi:10.1111/j.1875-595X.2003.tb00918.x
15. Luke Yeo IS. Dental implants: enhancing biological response through surface modifications. Dent Clin North Am. 2022;66(4):627–642. doi:10.1016/j.cden.2022.05.009)
16. Albrektsson T, Jemt T, Mölne J, Tengvall P, Wennerberg A. On inflammation-immunological balance theory-A critical apprehension of disease concepts around implants: mucositis and marginal bone loss may represent normal conditions and not necessarily a state of disease. Clin Implant Dent Relat Res. 2019;21(1):183–189. doi:10.1111/cid.12711
17. Choi JY, Albrektsson T, Jeon YJ, Yeo SL. Osteogenic cell behavior on titanium surfaces in hard tissue. J Clin Med. 2019;8:604. doi:10.3390/jcm8050604
18. Andrukhov O, Huber R, Shi B, et al. Proliferation, behavior, and differentiation of osteoblasts on surfaces of different microroughness. Dent Mater. 2016;32(11):1374–1384. doi:10.1016/j.dental.2016.08.217
19. Duarte PM, Cesar Neto JB, Goncalves PF, et al. Estrogen deficiency affects bone healing around titanium implants: a histometric study in rats. Implant Dent. 2003;12:340. doi:10.1097/01.ID.0000099750.26582.4B
20. Yang J, Farnell D, Devlin H, et al. The effect of ovariectomy on mandibular cortical thickness in the rat. J Dent. 2005;33:123. doi:10.1016/j.jdent.2004.09.001
21. Shirota T, Shirota T, Yamazaki M, et al. Bone mineral density in mandibles of ovariectomized rabbits. Clin Oral Implants Res. 2001;12:604. doi:10.1034/j.1600-0501.2001.120608.x
22. Jonasson G, Bankvall G, Kiliaridis S. Estimation of skeletal bone mineral density by means of the trabecular pattern of the alveolar bone, its interdental thickness, and the bone mass of the mandible. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001;92:346. doi:10.1067/moe.2001.116494
23. Yoshinari M, Oda Y, Inoue T, Matsuzaka K, Shimono M. Bone response to calcium phosphate-coated and bisphosphonate-immobilized titanium implants. Biomaterials. 2002;23:2879–2885. doi:10.1016/S0142-9612(01)00415-X
24. Astrand J, Aspenberg P. Topical, single dose bisphosphonate treatment reduced bone resorption in a rat model for prosthetic loosening. J Orthop Res. 2004;22:244–249. doi:10.1016/j.orthres.2003.08.008
25. Skoglund B, Holmertz J, Aspenberg P. Systemic and local ibandronate enhance screw fixation. J Orthop Res. 2004;22:1108–1113. doi:10.1016/j.orthres.2003.12.015
26. Tengvall P, Skoglund B, Askendal A, Aspenberg P. Surface immobilized bisphosphonate improves stainless-steel screw fixation in rats. Biomaterials. 2004;25:2133–2138. doi:10.1016/j.biomaterials.2003.08.049
27. Im JH, Kim SG, Oh JS, Lim SC. A comparative study of stability after the installation of 2 different surface types of implants in the maxillae of dogs. Implant Dent. 2015;24(5):586–591. doi:10.1097/ID.0000000000000292
28. Lazzara RJ, Testori T, Trisi P, Porter SS, Weinstein RL. A human histologic analysis of Osseotite and machined surfaces using implants with 2 opposing surfaces. Int J Periodontics Restorative Dent. 1999;19(2):117–129.
29. Hayakawa T, Yoshinari M, Nemoto K, Wolke JGC, Jansen JA. Effect of surface roughness and calcium phosphate coating on the implant/bone response. Clin Oral Implants Res. 2000;11(4):296–304. doi:10.1034/j.1600-0501.2000.011004296.x
30. Coelho PG, Granato R, Marin C. The effect of different implant macrogeometries and surface treatment in early biomechanical fixation: an experimental study in dogs. J Mech Behav Biomed Mater. 2011;4:1974–1981. doi:10.1016/j.jmbbm.2011.06.016
31. Duarte PM, de Vasconcelos Gurgel BC, Sallum AW, et al. Alendronate therapy may be effective in the prevention of bone loss around titanium implants inserted in estrogen-deficient rats. J Periodontol. 2005;76:107. doi:10.1902/jop.2005.76.1.107
32. Giro G, Sakakura CE, Gonçalves D, et al. Effect of 17β-Estradiol and Alendronate on the removal torque of osseointegrated titanium implants in ovariectomized rats. J Periodontol. 2007;78:1316. doi:10.1902/jop.2007.060390
33. Peter B, Pioletti DP, Laïb S, et al. Calcium phosphate drug delivery system: influence of local zoledronate release on bone implant osteointegration. Bone. 2005;36(1):52–60. doi:10.1016/j.bone.2004.10.004
34. Moreau M, Guillet C, Massin P, et al. Comparative effects of five bisphosphonates on apoptosis of macrophage cells in vitro. Biochem Pharmacol. 2007;73:718–723. doi:10.1016/j.bcp.2006.09.031
35. Jakobsen T, Baas J, Kold S, et al. Local bisphosphonate treatment increases fixation of hydroxyapatite-coated implants inserted with bone compaction. J Orthop Res. 2009;27:189–194. doi:10.1002/jor.20745
36. Cuaira´n C, Campbell PM, Kontogiorgos E, et al. Local application of zoledronate enhances miniscrew implant stability in dogs. Am J Orthod Dentofacial Orthop. 2014;145:737–749. doi:10.1016/j.ajodo.2014.01.020
37. Abtahi J, Tengvall P, Aspenberg P. Bisphosphonate coating might improve fixation of dental implants in the maxilla: a pilot study. Int J Oral Maxillofac Surg. 2010;39:673–677. doi:10.1016/j.ijom.2010.04.002
38. Sokmen N, Dundar S, Bozoglan A, et al. Effect of primary stabilisation on osseointegration of implants with local and systemic zoledronic acid application. J Craniofac Surg. 2022;33(5):1276–1281. doi:10.1097/SCS.0000000000008236
39. Abtahi J, Henefalk G, Aspenberg P. Impact of a zoledronate coating on early post-surgical implant stability and marginal bone resorption in the maxilla-A split-mouth randomized clinical trial. Clin Oral Implants Res. 2019;30(1):49–58. doi:10.1111/clr.13391
40. Zuffetti F, Testori T, Capelli M, Rossi MC, Del Fabbro M. The topical administration of bisphosphonates in implant surgery: a randomized split-mouth prospective study with a follow-up up to 5 years. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e168–76. doi:10.1111/cid.12151
41. Fiorillo L, Cicciù M, Tözüm TF, D’Amico C, Oteri G, Cervino G. Impact of bisphosphonate drugs on dental implant healing and peri-implant hard and soft tissues: a systematic review. BMC Oral Health. 2022;22(1):291. doi:10.1186/s12903-022-02330-y